eDoubler
Ivium Technologies
- Adds an extra applied/measured potential range to those instruments with higher compliance voltage
Product description
The eDoubler module is used to add an extra applied and measured voltage range to select instruments which have a higher compliance voltage. It is a module that is directly connected between the instrument and the cell cable.
It is compatible with the Vertex.One, dModule.1A/20V, CompactStat2.h20250, and XP20 instruments. In these cases the applied/measured voltage range is increased to match the compliance voltage, i.e. ±20V.
- Maximum applied voltage: ±20V
- Applied/measured resolution: Twice (2x) the primary potentiostat resolution
- Applied/measured accuracy: Twice (2x) the primary potentiostat error
- Input impedance: 100GOhm
- Electrometer input current: <150pA
- Electrometer bandwidth: Half (0.5x) the primary potentiostat bandwidth
- EIS frequency range: 10uHz to 50kHz (accurate without recalibration)