PalmSens  |  SKU: C-RM16CH-R-E-G1-ES4.F1

RackMount 16CH

Precio normal €0,00

EIS
Galvanic Isolation
Monunting type

Entrega y envío a la UE

Añadiremos en el presupuesto los gastos de envío, seguro y despacho de aduanas.

RackMount 16CH

PalmSens

16-channel potentiostat / galvanostat / EIS analyzer

  • Desktop or 19 inch rack-ready housing
  • Embedded Lablink Hub
  • FRA / EIS: 10 µHz up to 200 kHz
  • Potential range: ±3 V
  • Max. current: ±30 mA

High-Throughput Networked Electrochemical Analyzer

The RackMount 16CH is a high-performance 16-channel potentiostat built on the industry-proven EmStat4 LR platform. Engineered for high-throughput electrochemical research, it delivers precise, simultaneous measurements across all channels without compromise. Equipped with an embedded Lablink controller, the RackMount 16CH manages, buffers, and Connect electrochemical cells or disposable electrodes

16 independent channels

Each quad-cell connector interfaces with four independent electrochemical cells. Every channel features its own dedicated potentiostat with discrete Reference Electrode, Working Electrode, Counter Electrode, and ground connections for total isolation and control. Galvanic Isolation across the channels and hardware synchronization allow for using multiple Working Electrodes in the same Sharing, buffering and storage

Embedded Lablink controller

The RackMount 16CH, powered by Lablink, controls your measurements internally and logs them directly to its onboard Solid State Drive (SSD).

  • Independent Operation: Experiments run flawlessly without a PC connection.
  • Internal Logging: Data is saved locally, eliminating the risk from software interruptions.
  • Seamless Recovery: Simply reconnect
Status information at a glance

Status indicators and display

Each channel of the RackMount 16CH has a status indicator LED, showing if channels are available or claimed by users or if a measurement is running. The RackMount 16CH display shows Lablink-related status information and the fan speed.

Instant-start sensing

Automatic droplet detection

Optimized for Screen-Printed Electrodes (SPEs), each channel can be configured via software to trigger measurements automatically upon droplet detection. This minimizes the delay between
sample deposition and data acquisition, ensuring you capture the most time-sensitive electrochemical responses. It can also help streamline high-Efficient wiring

 

Adapter with shared Reference and Counter electrodes

Every RackMount 16CH includes four cell-connector adapters as standard. These DB25 interfaces internally bridge the Reference (RE) and Counter (CE) electrodes for each four-channel bank, eliminating the need to manually stack leads at the cell.


Voltammetric techniques

  •  Linear Sweep Voltammetry (LSV)
  • Cyclic Voltammetry (CV)
  • Fast Cyclic Voltammetry (FCV)
  • AC Voltammetry (ACV)

Potentiometric techniques

  • Linear Sweep Potentiometry (LSP)
  • Chronopotentiometry (CP)
  • MultiStep Potentiometry (MSP)
  • Open Circuit Potentiometry (OCP)

Pulsed techniques

  • Differential Pulse Voltammetry (DPV)
  • Square Wave Voltammetry (SWV)
  • Normal Pulse Voltammetry (NPV)

Impedimetric techniques

  •  Potentiostatic Electrochemical Impedance Spectroscopy (PEIS)
  • Galvanostatic Electrochemical Impedance Spectroscopy (GEIS)

Amperometric techniques

  • Chronoamperometry (CA) 
  • Zero Resistance Amperometry (ZRA)
  • Chronocoulometry (CC)
  • MultiStep Amperometry (MA)
  • Fast Amperometry (FAM)
  • Pulsed Amperometric Detection (PAD)

Other

  • Mixed Mode (MM)
  • Custom techniques (MethodSCRIPT)

Specifications

 

General
dc-potential range



±3 V
compliance voltage



±5 V
maximum current ±30 mA
max. data acquisition rate



1M samples /s


Potentiostat
applied potential resolution 100 µV
applied potential accuracy



≤ 0.2% ±1 mV offset
current ranges



1 nA to 10 mA
8 ranges
measured current resolution



0.009% of 

CR



(92 fA on 1 nA range)
measured 



< 0.2% of current
current accuracy

±20 pA  ±0.2% of range
bandwidth settings

320 Hz, 3.2 kHz, 30 kHz or 570 kHz


Galvanostat
current ranges



10 nA, 1 uA, 100 uA, 10 mA
4 ranges
applied dc-current ±3 * 

CR



(current range)
applied dc-current resolution 0.01% of CR
applied dc-



<0.4% of current
current accuracy

±20 pA  ±0.2% of range
potential ranges 50 mV, 100 mV, 200 mV, 500 mV, 1 V
measured dc-potential resolution 96 µV at ±3 V (1 V range)
48 µV at ±1.5 V (500 mV)
19.2 µV at ±0.6 V (200 mV)
9.6 µV at ±0.3 V (100 mV)
4.8 µV at ±0.150 V (50 mV)
measured dc-potential accuracy ≤ 0.2% ±1 mV offset
bandwidth settings

320 Hz, 3.2 kHz, 30 kHz or 570 kHz

 

FRA / EIS
frequency range 10 µHz to 200 kHz
ac-amplitude range 1 mV to 900 mV rms, or 2.5 V p-p

 

GEIS
frequency range 10 µHz to 100 kHz
ac-amplitude range 0.9 *CR (Amps)

 

Electrometer
electrometer amplifier input



> 1 TΩ // 10 pF
bandwidth



10 kHz default or
500 kHz for EIS and fast CA/CP


Other
electrode connections 16 x [ WE, RE, CE, and ground, with 2 mm banana plugs ]
housing Steel 19” housing:
44.4 x 33.5 x 4.4 cm (without front handles and feet)
extra height with feet: 1.5 cm
extra depth with handles: 4.0 cm
weight 4.5 kg
communication ethernet, I2C for peripheral control
Power 100-240 V, 90 W, 50-60 Hz
internal storage space 500 MB, equivalent to > 15M datapoints or ~1000 measurement files per channel
plus 512 GB SSD via internal Lablink module for entire instrument

 

Note
The accuracy contour plots were determined with an ac-amplitude of ≤10 mV rms for all limits, except for the high impedance limit, which was determined using an ac-amplitude of 250 mV. The standard cables were used. Please note that the true limits of an impedance measurement are influenced by all components in the system, e.g. connections, the environment, and the cell.

Connector options

We offer flexibility with a variety of connection options

There are four sensor connection options available, each supporting connection to four channels. When you add the RackMount16 to your basket, four units of your selected connection option will be included, providing connectivity for all 16 channels of the RackMount 16CH

Option A (recommended): Double shielded sensor cables with 2 mm banana connectors, compatible with standard croc clips. We include a set of croc clips for your convenience. If you would like to remove it from your quotation, just leave a comment in the quotation form.


Option B
: High-density cable that terminates in ferrules.

 

Option C: Connection (Screw) Terminal  to assemble your own wires.

 

Option D: Screen Printed Electrode Connector. We add a SPE Dummy Cell for your convenience. If you would like to remove it from your quotation, just leave a comment in the quotation form.

 

 

Brochoure.PDF

Manuals.PDF