Novonix  |  SKU: TCB-PHL-23- 56

Thermal Chambers 4-position, 20A

€0,00


Consegna e spedizione nell'UE

Nel preventivo aggiungeremo i costi di spedizione, assicurazione e sdoganamento.

Thermal Chambers 4-position, 20A

Novonix

Thermal Chambers

Stable and accurate temperature control

Accurate and stable temperature control of cells is critical for reducing noise in measured cell data. Designed specifically for UHPC, NOVONIX™ thermal chambers provide accurate and stable temperature control. They can operate from -20°C to +60°C and can accommodate a range of cell sizes. Thermal chambers are optimized for efficiency, keeping the cost down compared to other products on the market. 

All NOVONIX Thermal Chambers are CE marked and certified to IEC 62368-1:2018 for safety and emissions. Where required, UL/CSA marks may be added.


    • Set Point Temperature Range:
      10°C to 60°C
    • Accuracy (Stability) at
      20°C:±0.5°C (±0.05°C)
    • Accuracy (Stability) at 0°C: N/A
    • Required Lab Temp Range (Stability): 18°C to 28°C (±1°C)
    • Heating Method: Resistive
    • Cooling Method: Peltier
    • Standard Rack Mountable (Size): Yes (10U)
    • External Depth, inches
      (mm):
      21.4 (546)
    • Internal Dimensions, inches (mm): 14.2 x 12.5 x 15 (360) x (317) x (381)
    Front View
    Internal View

*Wide Range Chamber. NOVONIX only recommends these chambers for applications that will truly demand cold temperatures.
For regular applications, use Peltier-style chambers, which are more efficient and responsive due to their solid-state coolers.

 

Datasheet.PDF