ACCU  |  SKU: ACU30FC-SC-2SCCM-30SLM

ACCU ACU30FC Differential Pressure Mass Flow Controller / Meter

Prezzo normale €0,00

Model

Consegna e spedizione nell'UE

Nel preventivo aggiungeremo i costi di spedizione, assicurazione e sdoganamento.

ACCU ACU30FC Differential Pressure Mass Flow Controller / Meter

ACCU

The ACCU ACU30FC series is a differential pressure mass flow controller and flow meter for gases. Gas passing through a laminar throttling element creates a pressure drop that the sensor converts into an electrical signal; the flow is then calculated with the Hagen–Poiseuille equation. Because the measurement does not rely on the gas's specific heat, there is no specific-heat error, and close to 100 gases can be selected on the instrument without factory recalibration.

A built-in PID loop drives an electromagnetic proportional valve for precise flow control. The ACU30FC also measures volumetric flow, pressure and temperature, and shows all values on a large touch screen or sends them to a control system.

Key features

  • No specific-heat error; real-time, high-precision readings
  • Nearly 100 selectable gases without factory recalibration
  • Multi-parameter measurement: mass flow, volume flow, pressure, temperature
  • Very fast response (50 ms) with DSP-based control
  • Compact body with large touch-screen display
  • USB Type-C and Bluetooth connectivity, plus RS232/485 (Modbus), Profibus and DeviceNet

Applications

Vacuum coating, solar energy, analytical instruments, petroleum and petrochemicals, gas production and distribution, environmental protection, semiconductors, coal and metallurgy.

DATASHEET DOWNLOAD LINK

Technical specifications
Flow range 2 SCCM – 3000 SLM, depending on model
Turndown Controller 50:1 / Meter 100:1
Accuracy ±0.5% F.S.
Linearity / Repeatability ±0.5% F.S. / ±0.2% F.S.
Stability <±0.1% F.S.
Response time 50 ms
Temperature coefficient ±0.015% F.S./°C
Working temperature -10 – 60 °C
Working pressure Differential 0.05 – 0.5 MPa
Max. pressure 1 MPa
Leak rate 1×10-9 Pa·m³/s
Body / seals Stainless steel / Viton, NBR, FFKM, EPDM
Connections VCR, LOK, Ø8, Ø10, Ø12, flange and others
Protection / mounting IP40 / horizontal
Electrical connection DB9, RJ11, 5.5×2.1 power jack
Digital interface RS232 / RS485 Modbus, Profibus, DeviceNet, USB-C, Bluetooth
Analog signals 0–5 V, 4–20 mA, 1–5 V
Power supply ±15 VDC or 24 VDC

Configuration options (range, display, signals, power supply, communication, seals, fittings) are selected in the order code. Contact us for a quotation specifying your gas and flow range.